Does the deformation during a test also extend outward?
2012/11/4 19:01:00
A: Yes, so the distance near an edge or an adjacent indent is critical. ASTM recommends the
distance from the center of an indentation and the nearest adjacent
indentation be 3 times the diameter of the indentation. The distance between
the center of an indentation to the nearest edge must be at least 2.5 times the diameter of the indentation.