The AUT700 series offer various practical measurement features such as Through coating measurement for gauging thickness of a
painted object without removing the non mental coating.
The ultrasonic thickness gauge Model AUT700 series are featured as cost-efficiency and designed to make accurate measurements on steel, cast iron, aluminum, red copper, zinc, quartz glass, polyethylene,
PVC, gray cast iron and nodular cast iron. Thickness measurements are made from one side of the material with no need to cut
the part.
The AUT700 series offer various practical measurement features such as Through coating measurement for gauging thickness
of a painted object without removing the non mental coating
• Automatic Probe Recognition that recognizes transducer types, Field Calibration of all materials Sound velocity range
509 up to 18699m/s
• 2000 readings memory • Data output to PC
• High-Low Limits setting •
4 digit LCD display with settings
• Backlight display for dark environment and adjustable.
Features:
• Through coating measurement for gauging thickness of a painted object without removing the non mental coating.
• Thin coating mode and thick coating mode
• Available for gauging various materials such as metal, glass, plastic, rubber and etc.
• High measurement accuracy and wide range gauge applications • Suitable for pipe gauging.
• Whole range transducer models are available for almost all kinds of application, including coarse grain materialgauging and high temperature applications (Max. temperature up to 300℃).
• Adaptive probe function built in: Patent technology of automatic probe recognition: Automatically recognizingfor probes allowing the gage to match with probes made from various probe manufacturers.
• Power on automatic testing function contributing to measurement accuracy increasing.
• Auto power off detection and customizing power off configuration.
• Probe Zero function, Sound Velocity Calibration function.
• 9 velocities built in and programmable.
• Several practical measurement modes available: Standard Mode, Scanning Mode, Difference Mode, Average Mode,Maximum limitation Alert Mode, High Temperature Mode(Should be matched with high temperature probe).
• 8 button keyboard designing for convenient shortcut operations for Probe-Zero function, One-Point and Two-Point Sound-Velocity-calibration, as well as numerical adjusting with direction key control.
• Automatic Data Storage function: group data storage mode, and storage capacity for each group optional.
• Large data storage capacity, Max. up to 2000 groups.
• Special function of AUT700: USB data interface communication, easily accomplishing data transferring with PC andoutputting data with “.txt”
• Metric and imperial free convertible.
Thickness Measurement Range:Notes: Thickness range depends on material transducer type, surface condition and
temperature.Measuring Rang(Through Coating):3 mm -20 mm
Model
AUT700
Measuring range
0.75 –500mm(steel) (depends on probe)
0.3~200mm(Glass)
4~80mm(High temperature), 3~50mm(Gray Iron)Measuring Range
(Through Coating):
3 mm -25 mm
Tolerance
±(0.5%H+0.03)mm
H: thickness of the tested object
Measuring range of steel pipes
Φ15mm×1.0mm (probe: 7.5MHz, Φ6mm)
Φ10mm×1.2mm (probe: 7.5MHz, Φ6mm)
Data output
USB
Display resolution
0.1/0.01 mm or 0.01/0.001 inch
Battery life
280 hours typical battery life,
100 hours continuous with backlight onSound speed
509~18699m/s
Measuring units
mm/ inch
Dimensions(mm)
136 mm×72 mm×20mm
176g( Including Batteries)
Power supply
AA batteries (2Pcs)
Storage capability
2000 readings memory
Optional transducers、
5M, 7M, 2M, High temperature
• Through coating measurement for gauging thickness of a painted object without removing the non mental coating.
• Thin coating mode and thick coating mode
• Available for gauging various materials such as metal, glass, plastic, rubber and etc.
• High measurement accuracy and wide range gauge applications • Suitable for pipe gauging.
• Whole range transducer models are available for almost all kinds of application, including coarse grain materialgauging and high temperature applications (Max. temperature up to 300℃).
AUT700 Standard Configuration:
• Gage One Set
• Probe One Set(5MHz, Φ10mm)
• Couplant One Bottle(30ml, Glycerin)
• Battery: 2 AA Alkaline Batteries
• Guide Manual
• Hermetic Plastic Kit (18cm X 15cm X10cm)
• USB Cable
• Software Disk.
Model
Application
Gauge Range
Diameter
Frequency
Operating Temperature
5M Φ10
General Used
0.8—260mm
10mm
5MHz
-10~+50℃
5M Φ6
Small Pipes
0.75—60mm
6mm
5MHz
-10~+50℃
2M PT-12
Penetrating(Through Coating)
3.0—50mm(cast iron)
12mm or22mm
2MHz
-10~+50℃
7.5MΦ10
High-Accuracy
0.65—100mm(steel)
10mm
7.5MHz
-10~+50℃
ZW5P
High temperature
4.0—80mm
12mm
5MHz
-10~-310℃